DESCRIPTION:Speakers: Amaia Villa Abaunza (UCM)\nAuthors: A. Villa-Abaunza
 \, P. Ibáñez\, A. Espinosa and JM Udías\n\n**Introduction.** As radiati
 on research and cancer biology reach new horizons\, the need for improved 
 pre-clinical methodology has become paramount. In this work\, we are inter
 ested in designing an x-ray based irradiator for small animals and cells\,
  which is able to deliver a dose rate of 1 Gy/min. To do this\, some measu
 rements and Monte Carlo (MC) simulations have been performed to characteri
 ze different elements of the device.\n\n**Materials and methods.** On the 
 one hand\, a detailed characterization of different x-ray generators has b
 een performed in order to select the optimal beam parameters for our irrad
 iator. Absolute dose measurements with EBT2 Gafchromic films [1] have been
  performed in 3 x-ray tubes [2-4]. Different voltages and beam intensities
  were studied. On the other hand\, a shielding study has been done. We cal
 culated the lead shielding required for our device so that the irradiation
 s could be done in a controlled area (\n\nhttps://indico.ific.uv.es/event/